Test Procedure For The Oxygen Index Test

2019-01-31 15:00
Measuring the oxygen index can detect the ignition performance of the product and improve the safety performance of the product. The following is the Oxygen Index Test Apparatus test procedure for the oxygen index test.
1、Igniting the sample-start the time immediately and observe the burning time and combustion characteristics of the sample.
      If the combustion is suspended, but spontaneous reignition within 1 second, continue to observe and timing. If the burning time or burning length of the specimen does not exceed the form requirements, the test is recorded as an "O" reaction and the burning length or time is recorded. If one of the two exceeds the provisions in the table, the flame is extinguished and the test of the "X" reaction is recorded. Also record the combustion characteristics of the material, such as: droplets, soot, charcoal, drift combustion, combustion, afterglow or other characteristics that need to be recorded. If there is no burning flame, flammability should be reported as required, or include the oxygen index at flameless combustion. 
 
2、Remove the sample, wipe the combustion tube and the ignition surface of the dirt, so that the temperature of the combustion tube back to room temperature or another for the room temperature combustion tube, the next test.
     If the sample is long enough, the sample can be reversed or the re-used burned part is cut off. But can not be used to calculate the oxygen concentration. According to the above method, a set of samples was tested using the specific conditions of the "small sample lift method" as the amount of change.

3、If the combustion behavior of the previous sample is "X", the oxygen concentration is reduced; if the previous sample has a combustion behavior of "O", the oxygen concentration is increased.
     This test is repeated until the secondary oxygen concentration below the volumetric fraction does not exceed 1.0% and is O; after the any reaction is carried out at any suitable step. The reaction of this set of oxygen concentrations is referred to as the initial oxygen concentration Initial oxygen concentration Ψ. The above steps are repeated and the "X" or "O" response corresponding to Ψ0 is recorded. Is the first value of the NL series.

Related News

2023-10-26How do you measure oxygen levels in an experiment?
2023-10-26What is ASTM D6413?
2023-10-26What is ASTM D 695 standard test method for compressive properties of rigid plas
2023-10-26What types of construction materials are covered by ASTM standards?
2023-10-25Differences in the testing procedures between ASTM D2863 and ASTM D2863 17
2023-10-25What is the test method for limiting oxygen index?
2023-10-24What is 50% stretch in fabric?
2023-10-24How do you test fabric quality?
2023-10-24How are the materials are tested as per ASTM standards?
2023-10-24Can the oxygen index test be used to compare the fire resistance properties of d

Copyright 2022:Qinsun Instruments Co., Limited

High-end textile tester supplier  Email:info@qinsun-lab.com | Textile Testing Equipment pdf | Tel:021-67800179 |