ASTM F1980: Accelerated Aging Time and Temperature

2023-10-31 13:50

ASTM F1980 is a standard guide provided by the American Society for Testing and Materials (ASTM) that outlines a recommended protocol for accelerated aging testing for medical devices. It provides guidance on determining the appropriate time and temperature parameters for accelerated aging tests. Here is a summary of the key points regarding accelerated aging time and temperature according to ASTM F1980:


1. Selection of reference aging time and temperature: The standard suggests selecting a reference aging time and temperature combination that represents the anticipated storage time and conditions of the medical device in real-world use. This serves as a benchmark for accelerated aging testing.


2. Arrhenius equation: The standard recommends using the Arrhenius equation to estimate the test duration at elevated temperatures. The equation considers the temperature sensitivity of the material and helps extrapolate the aging behavior at accelerated temperatures.


3. Acceleration factor (AF): The acceleration factor is calculated using the Arrhenius equation and represents how much faster the aging process occurs at the elevated test temperature compared to the reference temperature. The AF is typically determined by industry-specific guidelines, regulatory requirements, or specific product standards.


4. Test duration: The standard suggests determining the accelerated aging test duration by dividing the anticipated product shelf life (in months) by the acceleration factor. This calculation provides an estimation of the test duration required to simulate the expected aging effects.


5. Rationale for test durations: The standard acknowledges that there may be deviations from the calculated test duration based on factors such as the criticality of the medical device, expected aging mechanisms, and available historical data. Therefore, it recommends using engineering judgment to establish appropriate test durations.


6. Continuous vs. intermittent testing: The standard specifies that accelerated aging tests may be performed continuously (constant test temperature) or intermittently (cyclic temperature exposure). The selection of continuous or intermittent testing should be based on the intended use and characteristics of the medical device.


It is essential to note that ASTM F1980 provides a general guideline, and specific product standards or regulations may have their own requirements for accelerated aging time and temperature. Therefore, it is recommended to consult the relevant standards, guidelines, or regulatory bodies specific to your industry when conducting accelerated aging testing for medical devices.



Related News

2023-10-31What is the ASTM standard for Accelerated Aging?
2023-10-31ASTM F1980: Accelerated Aging Time and Temperature
2023-10-31How do you do accelerated Aging testing?
2023-10-31What are the two types of tensile testing machines?
2023-10-30What is a Universal Tensile Testing Machine and how does it work?
2023-10-30How to Conduct Precise Material Strength Testing with a Universal Tensile Testin
2023-10-30What types of textiles are covered by textile standards?
2023-10-30What is ASTM textile standard?
2023-10-27What are the standards for accelerated aging?
2023-10-27What is the purpose of the Montech ABR 3000 abrasion tester?

Copyright 2022:Qinsun Instruments Co., Limited

High-end textile tester supplier  Email:info@qinsun-lab.com | Textile Testing Equipment pdf | Tel:021-67800179 |