Site: Home > Home > Products > Application > Mask & Protective Clothing Tester

Masks Bacterial Filtration Efficiency (BFE)Tester

Description:Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

Factory Direct SalesQuality AssuranceLightning DeliveryHassle-free Service
Free Quote
Email
Masks Bacterial Filtration Efficiency (BFE)Tester

Product Details

Mask Bacterial Filtration Efficiency Tester is applied to test the filtration efficiency of mask, can be used in metering calibration department, scientific research institution, mask manufacturer and other related departments.
 
Standards:
YY 0469-2004 Technical requirements for surgical mask
BS EN 14683-2014 Medical face masks - Requirements and test methods
ASTM F2100-2019 Standard Specification for Performance of Materials Used in Medical Face Masks            
ASTM F2101 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
 
Feature:
1.Bacterial Filtration Efficiency Tester Of Mask: Negative pressure testing system, which can protect operator’s safety.
2.Peristaltic pump was built in the negative pressure cabinet, A & B two routes 6 level Andersen sampling head.
3.The flow speed of peristaltic pump can be set.
4.The spray flow of bacterial liquid inside the special microbial aerosol generator can be set.
5.BFE tester good atomization effect.
6.Embedded high speed industrial microcomputer control.
7.10.4 "industrial high brightness color touch screen display.
8.USD connector, support USB data archived
9.Mask BFE Tester high brightness light installed in the cabinet.
10.Switching type glass door in front of the machine, easy to observe by the operator.
 
Specification:
Key Specification Specification Range Resolution Max. permissible errors
A Route Sampling Flow 28.3L/min 0.1L/min Within ± 2.0%
B Route Sampling Flow 28.3L/min 0.1L/min Within ±2.0%
Spray Flow (8~10)L/min 0.1L/min Within ±5.0%
Peristaltic Pump Flow (0.006~3.0)mL/min 0.001ml/min Within ±2.0%
Front Pressure of A Route Sampling Flowmeter (-20~0)kPa 0.01kPa Within ±2.0%
Front Pressure of B Route Sampling Flowmeter (-20~0)kPa 0.01kPa Within ±2.0%
Front Pressure of Spray  Flowmeter (0~300)kPa 0.1kPa Within ±2.0%
Negative Pressure of the Chamber (-90~-120)Pa 0.1Pa Within ±1.0%
Working Temperature 0-50℃
Negative Pressure of the Chamber >120Pa
Data Save Ability >500000 sets(Scalable capacity)
High Efficiency Particulate Air Filter Properties ≥99.995%@0.3μm, ≥99.995%@0.12μm
Median diameter of aerosol generator mass Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5
Double Routes 6-lever Andersen Sampling apparatus LevelⅠ>7μm;          Level Ⅱ: (4.7~7)μm;
Level Ⅲ: (3.3~4.7)μm;  Level Ⅳ: (2.1~3.3)μm;
Level Ⅴ: (1.1~2.1)μm;  Level Ⅵ: (0.6~1.1)μm
Size of Aerosol Chamber 600×80×3mm (Length×Diameter×Thickness)
Total number of positive quality control sampler particles (2200±500)cfu
Negative Chamber Flow Speed ≥5m3/min
Size of Main Machine

1470×790×2100mm (L×W×H)
 
Power Supply AC220V±10%, 50Hz
Noise of the Tester <65dB(A)
Weight ≈150kg
Power Consumption <1500W

Mask Bacterial Filtration Efficiency Tester

Related Products

  • Tensile Strength Testing Equipment for Lea

    Tensile Strength Testing Equipment for Lea

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • Taber Leather Abrasion Resistance Tester

    Taber Leather Abrasion Resistance Tester

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • IEC60695 Glow Wire IEC Fire Hazard Testing

    IEC60695 Glow Wire IEC Fire Hazard Testing

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • ASTM D 2863, ISO 4589-2 Automatic Limited

    ASTM D 2863, ISO 4589-2 Automatic Limited

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • ISO105B02 Standard Textiles Xenon Aging La

    ISO105B02 Standard Textiles Xenon Aging La

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • Bag Zipper Reciprocation Pull Fatigue Test

    Bag Zipper Reciprocation Pull Fatigue Test

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • ISO 6925 Carpet Flammability Testing Equip

    ISO 6925 Carpet Flammability Testing Equip

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

  • Portable Mask Breathing Resistance Testing

    Portable Mask Breathing Resistance Testing

    Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

Leave Message

Send message or email info@qinsun-lab.com get favorable prices,reply tu you within 24 hours!

Copyright 2022:Qinsun Instruments Co., Limited

High-end textile tester supplier  Email:info@qinsun-lab.com | Textile Testing Equipment pdf | Tel:021-67800179 |